Electrochemical glucose biosensor based on nickel oxide nanoparticle-modified carbon paste electrode.
Ceren Erdem, Derya Koyuncu Zeybek, Gözde Aydoğdu, Bülent Zeybek, Sule Pekyardımcı, Esma Kılıç
Index: Artif. Cells. Nanomed. Biotechnol. 42(4) , 237-44, (2014)
Full Text: HTML
Abstract
In the present work, we designed an amperometric glucose biosensor based on nickel oxide nanoparticles (NiONPs)-modified carbon paste electrode. The biosensor was prepared by incorporation of glucose oxidase and NiONPs into a carbon paste matrix. It showed good analytical performances such as high sensitivity (367 μA mmolL(-1)) and a wide linear response from 1.9×10(-3) mmolL(-1) to 15.0 mmolL(-1) with a limit of detection (0.11 μmolL(-1)). The biosensor was used for the determination of glucose in human serum samples. The results illustrate that NiONPs have enormous potential in the construction of biosensor for determination of glucose.